The DS500 is an in-situ water quality sonde based on advanced sensor technology. It can be used portably in the field or deployed long-term for online monitoring of various parameters, including temperature, depth, conductivity (specific conductivity, salinity, TDS, resistivity, density), pH, ORP (oxidation-reduction potential), dissolved oxygen, turbidity, chlorophyll, blue-green algae, oil in water, ammonium ions, and more. With a compact diameter and a wide range of measurement parameters, it is widely applicable in various water quality testing scenarios, including marine, surface water/groundwater, aquaculture, environmental protection, and water supply and drainage.

Product Features

Titanium alloy material for higher hardness, lighter weight, and better corrosion resistance.

Replaceable battery design with indicator light, especially convenient for field operations.

Single cable multi-host cascading allows simultaneous measurement of more water quality parameters or those at different water profiles.

The same instrument has measurement/storage/calibration modes.

Bluetooth connection to a mobile app (with accompanying app).

Equipped with a low-flow sampling system for groundwater and profile monitoring sampling functions.

High reliability testing and evaluation system, stable measurements, and strong anti-interference capability.

Uses RS485 interface and standard Modbus RTU protocol for real-time data transmission.

Strong expandability, with interchangeable sensors that can be freely combined and easily replaced, plug-and-play, allowing multiple sensors to be installed simultaneously.

Customizable high-strength cables with quick-release twist-lock connectors for efficient field operations.

Electric cleaning brush ensures the measurement window is clean, preventing microbial growth and reducing labor costs.

Compact structure with strong portability, minimum diameter of 49 mm, suitable for installation in smaller spaces.

Wide application range, suitable for portable field use or long-term deployment for online monitoring.



Application Fields

Technical Specifications

Sensor

Measurement Method

Applicable water depth

Measurement Range

Accuracy

Resolution

Display Units

Temperature

Thermistor Method

0-100 m

-5 to +70°C

±0.1°C

0.01°C

Depth

Piezoresistor Method

0-100 m

0-100 m (can be 20 m, 20 m, 30 m)

±0.25%F.S

1mm

μs/cm,mS/cm

Conductivity/Temperature

Four-electrode Method

0-100 m

0-200,000 µS/cm; 0-300,000 µS/cm(typical value)

±0.5%  of reading or ±1μS/cm

0.01μS/cm

μS/cm,mS/cm

Salinity

Export/Calculate Parameters

0-180PSU

±1% or ±0.1, whichever is greater

0.01ppt

ppt,PSU

Total Dissolved Solids

0-130ppt

 ±1% of reading

0.01ppt

ppt,ppm

Density

-

 ±1% of reading

-

g/L,kg/L

Resistivity

0-2000,000Ωcm

±1% of reading

0.01Ωcm

MΩcm,KΩcm,Ωcm

Dissolved Oxygen/Temperature

Fluorescence Quenching

Method

0-50m

RDO Concentration0-50 mg/L

0-20mg/L:±0.2mg/L

20-50mg/L:±5%(typical value)

0.01mg/L

mg/L

0-50m

RDO Saturation:0-500%

0-200%:±1%

2005-500%:±5%(typical value)

0.01%

%

Turbidity

ISO-7027 Method

0-100m

0-4000 NTU

0-10NTU:±0.1 NTU

10-100NTU:±2% of reading or ±1NTU

100-4000NTU:±5% of reading

0.01NTU

NTU,FNU,ppt

Total Suspended Solids

Export/Calculate Parameters

0-1500mg/L

±5% of reading

0.1mg/L

mg/L,ppt

pH

Electrode method

0-50m

0-14

±0.1

0.01

-

ORP

Electrode Method

0-50m

-200 ~ +2000 mV

±5 mV (standard solution)

0.01mV

mV

Chlorophyll a

Fluorescence Method

0-100m

0-500 μg/L

Serial dilution of chlorophyll a equivalent concentration from 0-500 μg/L of Rhodamine WT solution, R² > 0.999.

0.01μg/L

μg/L,cell/L

(Blue-green algae)
phycoerythrin

Fluorescence Method

0-100m

0-100 µg/L

Serial dilution of Rhodamine WT solution with an equivalent concentration of 100μg/L PE, R² > 0.999

0.01μg/L

μg/L,cell/L

(Blue-green algae) phycocyanin

Fluorescence Method

0-100m

0-100 µg/L

Serial dilution of Rhodamine WT solution with an equivalent concentration of 100μg/L PC, R² > 0.999.

0.01μg/L

μg/L,cell/L

Suspended Solids

Backscatter 45°

0-100m

0-1000 mg/L

0-100mg/L:0.1mg/L or  ±2% ,whichever is greater

100-1000mg/L:  ±5% of reading

0.01mg/L

mg/L,NTU

fDOM

Fluorescence Method

0-100m

0-500 ppm

Serial dilution of 0 - 200 ppm quinine sulfate solution, R² > 0.99

0.01ppm

ppm

COD

UV Method

0-100m

0.1-200 mg/L (equiv. KHP)

±2% of reading or 0.5mg/L (equiv. KHP), whichever is greater

0.01mg/L (equiv. KHP)

mg/L

Oil in water

Fluorescence Method

0-100m

0-200 ppm

 ±3% full scale

0.01ppm

ppm

Silver Ion

Ion Electrode Method

0-15m

0-100,000 ppm

±10% of reading  or 2mg/L, whichever is greater

0.01ppm

mg/L,ppm

Sulfide Ion

Ion Electrode Method

0-15m

0-30,000 ppm

±10% of reading or 2mg/L, whichever is greater

0.01ppm

mg/L,ppm

Fluoride Ion

Ion Electrode Method

0-15m

0-1000 ppm

±10% of reading or 2mg/L, whichever is greater

0.01ppm

mg/L,ppm

Ammonium Ion

Ion Electrode Method

0-15m

0-1000 ppm

±10% of reading or 2mg/L, whichever is greater

0.01ppm

mg/L,ppm

Ammonia Nitrogen

Export/Calculate Parameters

0-15m

0-200mg/L

±10% of reading or 2mg/L, whichever is greater

0.01ppm

mg/L,ppm

Sulfate Ion

Ion Electrode Method

0-15m

0-1000ppm

±10% of reading or 2mg/L, whichever is greater

0.01ppm

mg/L,ppm

Bromide Ion

Ion Electrode Method

0-15m

0-1000ppm

±10% of reading or 2mg/L, whichever is greater

0.01ppm

mg/L,ppm

Iodide Ion

Ion Electrode Method

0-15m

0-1000ppm

±10% of reading or 2mg/L, whichever is greater

0.01ppm

mg/L,ppm

Cadmium Ion

Ion Electrode Method

0-15m

0-10,000ppm

±10% of reading or 2mg/L, whichever is greater

0.01ppm

mg/L,ppm

Lead Ion

Ion Electrode Method

0-15m

0-20,000ppm

±15% of reading or 5mg/L, whichever is greater

0.01ppm

mg/L,ppm

Chloride Ion

Ion Electrode Method

0-15m

0-1000ppm

±10% of reading or 2mg/L, whichever is greater

0.01ppm

mg/L,ppm

Potassium Ion

Ion Electrode Method

0-15m

0-1000ppm

±10% of reading or 2mg/L, whichever is greater

0.01ppm

mg/L,ppm

Copper Ion

Ion Electrode Method

0-15m

0-5000ppm

±10% of reading or 2mg/L, whichever is greater

0.01ppm

mg/L,ppm

Calcium Ion

Ion Electrode Method

0-15m

0-1000ppm

±10% of reading or 2mg/L, whichever is greater

0.01ppm

mg/L,ppm

Cyanide Ion

Ion Electrode Method

0-15m

0-250ppm

±10% of reading or 2mg/L, whichever is greater

0.01ppm

mg/L,ppm


Product Selection

Model

Outer Dimensions

Weight

Operating Temperature

Storage Temperature

Waterproof Rating

Communication Protocol

Power Supply Voltage

Plug and play Sensor Slots

Power Consumption

DS500

Diameter 4.9 cm, Length 49 cm

 

1.25 kg

 -5 to +70°C

 

-40 to +70°C

 IP68

 

RS485,Bluetooth

 

12-24V

 

4

Measurement:60 mA typical,120 mA  Max

DS501

Diameter 5,7cm, Length 49 cm

1.50kg

 -5 to +70°C

 

-40 to +70°C

 IP68

 

RS485,Bluetooth

 

12-24V

 

5

Measurement:65 mA typical,120 mA  Max

DS502

Diameter 7.5 cm, Length 49 cm

2.30kg

 -5to+70°C

 

-40 to +70°C

 IP68

 

RS485,Bluetooth

 

12-24V

 

6

Measurement:80 mA typical,120 mA  Max